UV Light Meter
Light Meter Features
- Less than 1% Probe to Probe Repeatability.
- Real Time Measurements.
- Single Channel probe
- Dual Channel Probe capability
- Interchangeable Probes
- Computer Interface
- Cosine response probes
- Measures mw/cm2, Peak mw/cm2, mj/cm2(SP), Accumulated mj/cm2 (total time-seconds) and Accumulated time (total seconds)
- Calculates average value of mw/cm2, calculates mw/cm2 % uniformity and % deviation from center value
- X10 range
- Battery operates for approximately 18 hours before recharge
- Charger operates from 100-240 VAC, 50-60 Hz with UL, CE, and TUV approvals
Comprehensive UV Light Measurement Instrument
The Model 325 built-in computer allows storing of sensor readings and calculations of average value and percent deviations. The two line display format allows for several readings at one time. The Model 325 measures MW/CM2, Peak mw/cm2, mw/cm2 and time (single pulse) with average mw/cm2 and Peak mw/cm2.
The Model 325 has four modes of operation. 5 point mw (mw/cm2) and 5 point mj (mj/cm2) allow taking five readings in a fixed pattern and then displaying the average value and the +/-% uniformity as well as the +/-% deviation from the center value.
The rechargeable battery lasts for about 18 hours and the charger operates from 100-240 VAC, 50-60 Hz with UL, CE, and TUV approvals.
A variety of sensor packages can be used with the Model 325 at wavelengths from 220nm to 540nm and beyond. Traceability to NIST is +/-3.0% with probe matching to +/-1.0%.
Light Meter Model 325 Specifications |
|||
Switch Functions: |
|
||
Display: |
|
||
Range: |
|
||
Mode Selection: | mw/cm22 |
|
|
mj/cm22 & mw/cm22 |
|
||
5 Point mw |
|
||
5 Point mj |
|
||
Display Pattern: | |||
Power: |
|
||
Probe Wavelengths: |
|
||
Size: |
|
||
Weight: |
|
This graph accurately represents the spectral response of probes used today in the Microelectronic Industry. This graph includes the spectral response of the detector as well as the spectral response of the filter glass used in probes.